749-0798
GC Analysis of Semivolatiles in Drinking Water on a 5% Phenyl Column after SPE using ENVI™-18 DSK
suitable for GC, application for SPE
About This Item
Prodotti consigliati
tecniche
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: acidified drinking water
SPE tube/cartridge: ENVI-18 DSK SPE Disk, 47 mm (57171)
condition: 5 mL dichloromethane; 5 mL methanol; 5 mL DI water
sample addition: 5 mL water sample
elution: 2 x 5 mL acetonitrile rinsing internal chamber of sample reservoir thoroughly
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C to 160 °C (3 min), then to 300 °C (3 min) at 6 °C/min
detector: MS, scan range m/z = 45-450
carrier gas: helium, 33 cm/sec
injection: 1 μL split/splitless, 1 min delay
Compatibilità
application for SPE
applicazioni
environmental
food and beverages
Applicazioni
Adjust 1 L drinking water to pH <2 with 6 N HCl. Add 5 mL methanol and mix thoroughly
Note legali
Analita
Colonna analitica
Prodotti correlati
Choose from one of the most recent versions:
Certificati d'analisi (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.