Accéder au contenu
Merck

TLC Analysis Of Plant Extracts on HPTLC Silica gel 60 NH2 F254s

TLC Analysis Of Plant Extracts on HPTLC Silica gel 60 NH2 F254s application for TLC

Matériaux

À utiliser avec

Réf. du produit
Description
Tarif

Vaporisateur de type flacon

capacity 240 mL

Cuve de développement de plaques CCM Aldrich® - rectangulaires

L × H × W 27.0 cm × 26.5 cm × 7.0 cm

Cuve de vaporisation pour CCM Sigma

Lampe UV pour Spectroline® Séries-E

output8 W, AC/DC input 115 V AC

Cuve de développement de plaques CCM Aldrich® - rectangulaires

L × H × W 7.5 cm × 15.5 cm × 8.0 cm

Seringues TLC Hamilton®

1702, volume 25 μL, needle size 22s ga, needle L 51 mm (2 in.),

coated tip

Lampe UV pour Spectroline® Séries-E

output6 W, AC/DC input 120 V AC

Seringues TLC Hamilton®

1705, volume 50 μL, needle size 22 ga, needle L 51 mm (2 in.),

coated tip

Seringues TLC Hamilton®

1701, volume 10 μL, needle size 26s ga, needle type, fixed, needle L 51 mm (2 in.),

coated tip

Cuve de développement de plaques CCM Aldrich® - rectangulaires

L × H × W 17.5 cm × 16.0 cm × 8.2 cm

Lampe UV pour Spectroline® Séries-E

output6 W, AC/DC input 230 V AC

Lampe UV pour Spectroline® Séries-E

output4 W, AC/DC input 230 V AC

Lampe UV pour Spectroline® Séries-E

output8 W, AC/DC input 230 V AC

Latch-lid TLC developing chambers

For use with 10 cm × 10 cm plates

Seringues TLC Hamilton®

1710, volume 100 μL, needle size 22 ga, needle L 51 mm (2 in.),

coated tip

Latch-lid TLC developing chambers

For use with 20 cm × 20 cm plates

Lampe UV pour Spectroline® Séries-E

output4 W, AC/DC input 115 V AC

Pipettes micro-capillaires Kimble®

Pipettes micro-capillaires Kimble®

Pipettes micro-capillaires Kimble®

capacity 20 μL, accuracy: ±0.5%

Reagent sprayer for TLC plates

volume 250 mL, Bottle-Type

CONDITIONS

sample preparation

2 g plant material extracted for 1 h at 60°C under shaking with 50 ml solvent. Filtration of crude extract with an 0.45 μm syringe filter. Application by CAMAG ATS4 6 mm bandwise.

plate

1.15647 HPTLC Silica gel 60 RP-18 F 254s MS-grade 20cm x 10 cm (1.151647.0001)

eluent

[A] ethylacetat; [B] formic acid; [C] glacial acid; [D] water; (100:11:11:27, A:B:C:D) containing 0.1% TFA

migration distance

6 cm

chamber

normal chamber without chamber saturation

sample

0.150 μL

detector

UV, 254 nm (Identification by coupling the TLC – MS Interface (CAMAG) with the Advion expression) (CMS mass spectrometer (+)-ESI mode. Eluent: Acetonitrile / Water 95:5 + 0.1% TFA.)