HPLC Analysis of Cinacalcet Hydrochloride Impurities with a Purospher® STAR Phenyl Column

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Purospher® STAR Phenyl (5 µm) LiChroCART® 150 x 4.6 mm I.D. (1.51922)
mobile phase: [A] 10 mmol/L Sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid; [B] Acetonitrile
gradient: 0-0.5 min 2.5% B, in 4.5 min to 95%B, 2 min 95%B, 3 min at 2.5%B
flow rate: 1.0 mL/min
pressure: 60 bar (870 psi)
column temp.: 60°C (Sampler: 10 °C)
detector: UV, 215 nm (analytical flow cell; 10 µL)
injection: 5 µL
sample/matrix: Impurity identification solution with cinacalcet impurities A, B, C at 1.5 µg/mL each and cinacalcet at 5 µg/mL in solvnt mixture (10 mmol/L sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid, and acetonitrile in a ratio of 50:50 (%v/v))

Application(s)

pharmaceutical

Description générale

Cinacalcet hydrochloride is an oral calcimimetic indicated for the treatment of secondary hyperthyroidism in patients on dialysis with end-stage renal disease, and in patients with parathyroid carcinoma to reduce hypercalcemia.

Remarque sur l'analyse

The developed method on a Purospher® STAR Phenyl, 150 × 4.6 mm (5µm) column was found to be linear for cinacalcet hydrochloride up to a concentration of 6 µg/mL with a LOQ of 0.65 µg/mL and LOD of 0.21 µg/mL. For cinacalcet impurity A, impurity B and impurity C, the method was linear up to a concentration of 1.8 µg/mL with LOQ ranging from 0.03 to 0.13 µg/mL and LOD ranging from 0.01 to 0.04 µg/mL. The method can be applied for the study of related substances in cinacalcet hydrochloride API using the Purospher® STAR Phenyl HPLC column. Repeatability for the cinacalcet 5 µg/mL solution was 1.19% RSD (n=5). The tailing factor for cinacalcet eluting at 12.3 min was NMT 2.0 (1.31) and the theoretical plate count was significantly above 2000.

Autres remarques

App_501

Informations légales

LICHROCART is a registered trademark of Merck KGaA, Darmstadt, Germany
Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Development and validation of a stability-indicating RP-UPLC method for the estimation of impurities in cinacalcet hydrochloride API and its formulation
Reddy PS, et al.
Scientia Pharmaceutica, 83(4) (2015)

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique