Direkt zum Inhalt
Merck

G1007615

LC/MS Analysis of Nucleotides on SeQuant® ZIC-cHILIC

application for LC-MS

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

Methode(n)

LC/MS: suitable

Testparameter

column: SeQuant® ZIC-cHILIC, 10 cm x 2.1 mm, 3.0 μm particles (1.50657)
mobile phase: [A] acetonitrile; [B] 50 mM ammonium acetate, pH 5.0 with acetic acid in water
gradient: 26 to 27% B in 10 min; to 27 to 35% B in 10 min; to 26% B in 0.1 min; hold at 26% B for 9.9 min
flow rate: 0.3 mL/min
pressure: 1029 psi (71 bar)
column temp.: 50 °C
detector: Q-TOF (ESI-negative), XIC of the following m/z values: AMP-346, ADP-426, ATP-506, CMP-322, CDP-402, (CTP-482, GMP-362, GDP-442, GTP-522, UMP-323, UDP-403, UTP-483)
injection: 10 μL
sample: 10-30 μg/mL of nucleotides (AMP, ADP, ATP, CMP, CDP, CTP, GMP, GDP, GTP, UMP, UDP, UTP) (80/20 acetonitrile/50 mM ammonium acetate in water, pH 5.3 with acetic acid)

Eignung

application for LC-MS

Anwendung(en)

pharmaceutical (small molecule)

Hinweis zur Analyse

Nucleotides are the building blocks of nucleic acids. They comprise a nitrogenous base, a sugar, and a phosphate group making them highly polar and ionic. This causes a challenge for typical reversed phase chromatography (RPC). These charged polar analytes are poorly retained by RPC, and thus typically chromatographed by either ion-exchange chromatography or RPC in conjunction with an ion-pairing reagent. Nucleotide analysis using ion-pair reagents or high concentration of phosphate buffers cause high background and ion source pollution for MS detection. HILIC separation is an alternative that permits sensitive MS detection and without the use of ion-pair reagents.

Rechtliche Hinweise

SeQuant is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyt

Beschreibung

    Adenosine 5′-monophosphate sodium salt, from yeast, ≥99%
    Adenosine 5′-diphosphate sodium salt, bacterial, ≥95% (HPLC)
    Adenosine 5′-triphosphate disodium salt hydrate, BioXtra, ≥99% (HPLC), from microbial
    Cytidine 5′-monophosphate, Sigma Grade, ≥99% (HPLC), synthetic, powder
    Cytidine 5′-diphosphate sodium salt hydrate, from yeast, ≥95%
    Cytidine 5′-triphosphate disodium salt, ~95%
    Guanosine 5′-monophosphate disodium salt hydrate, from yeast, ≥99%
    Guanosine 5′-diphosphate sodium salt, Type I, ≥96% (HPLC)
    Guanosine 5′-triphosphate sodium salt hydrate, ≥95% (HPLC), powder
    Uridine 5′-monophosphate, ≥98%
    Uridine 5′-(trihydrogen diphosphate) sodium salt from Saccharomyces cerevisiae, 95-100%
    Uridine 5′-triphosphate trisodium salt hydrate, from yeast, Type III, ≥96%
    Alle anzeigen (12)

Choose from one of the most recent versions:

Analysenzertifikate (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Kundensupport

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.