Direkt zum Inhalt
Merck

G005445

HPLC Analysis of Milk Thistle Standards on Ascentis® Express C18

application for HPLC

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

Methode(n)

HPLC: suitable

Testparameter

column: Ascentis Express C18, 10 cm x 3 mm I.D., 2.7 μm particles (53814-U)
mobile phase: [A] water with 0.1% formic acid; [B] methanol
gradient: held at 35% B for 3 min; 35 to 45% B in 10 min; held at 45% B for 2 min; 45 to 100% B in 5 min
flow rate: 0.6 mL/min
pressure: 4105 psi (283 bar)
column temp.: 35 °C
detector: UV, 254 nm
injection: 20 μL
sample: 20 μg/mL in 80:20, water:methanol

Eignung

application for HPLC

Anwendung(en)

vitamins, nutraceuticals, and natural products

Hinweis zur Analyse

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of milk thistle related compounds.

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Choose from one of the most recent versions:

Analysenzertifikate (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Kundensupport

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.