Direkt zum Inhalt
Merck

CR0102006

Millipore

Polygard® CR Kartuschenfilter

pore size 1.0 μm, cartridge nominal length 20 in. (50 cm), Code 0 (2-222; O-rings)

Synonym(e):

Polygard-CR Cartridge Filter 20 in. 1.0 μm Code 0 Silicone

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

UNSPSC-Code:
23151806
eCl@ss:
32031690

Materialien

polypropylene
silicone seal

Qualitätsniveau

Leistungsmerkmale

hydrophilic

Hersteller/Markenname

Polygard

Parameter

4.8 bar max. differential pressure (70 psid) at 25 °C
4.8 bar max. inlet pressure (70 psi) at 25 °C

Nennlänge Kartusche

20 in. (50 cm)

Durchmesser

7.0 cm (2.75 in.)

Verunreinigungen

≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)

Gravimetrisch bestimmte extrahierbare Substanzen

≤50 mg

Porengröße

1.0 μm nominal pore size
1.0 μm pore size

Aufnahme

sample type liquid

Kartuschencode

Code 0 (2-222; O-rings)

Allgemeine Beschreibung

The Polygard® CR Cartridge Filter products are nominally rated filters designed for particle removal applications in liquids and gases. The graded-density depth structure of Polygard-CR filter media provides maximum filtration capacity, a 1.0 μm pore size and the all-polypropylene construction offers low extractables levels and broad chemical compatibility.

Polygard-CR cartridge filters provide high throughput with 4.8 bar max. differential pressure (70 psid) at 25 °C.

Typical Applications
Polygard-CR high-efficiency cartridge filters, composed of graded density polypropylene media are widely used in the beverage industry for liquid and gas filtration applications to protect tighter filters.

Regulatory Compliance
Polygard® CR Cartridge Filters are designed, developed, and manufactured in accordance with a Quality Management System approved by an accredited registering body to an ISO 9000 Quality Systems Standard and are shipped with a Certificate of Quality. Each cartridge filter is supported by a Validation Guide for compliance with regulatory requirements.

For traceability and easy identification, each filter is marked with identifying characteristics.

Multiple Formats Available
Polygard-CR product line offers a full range of filtration areas is available to suit varying application requirements. A variety of connection options are offered for easy adaptation to existing housings.

These devices have been developed to minimize hold-up volume when evaluating multiple media configurations.

Angaben zur Herstellung

Sanitization Protocol
30 min mit 80 °C heißem Wasser
Sterilization Method
10 Autoklaven- oder In-situ-Dampfsterilisationszyklen von 30 min bei 126 °C

Hinweis zur Analyse

Erfüllt die Anforderungen des USP-Tests auf oxidierbare Substanzen nach Durchspülen mit ≥10 l Wasser
Gravimetric Extractables: after 24-h soak in water

Rechtliche Hinweise

POLYGARD is a registered trademark of Merck KGaA, Darmstadt, Germany

Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.