Most applications will require pre-treatment prior to dialysis. As mentioned in the 'DESCRIPTION' section, under 'Preparation Note', for the removal of glycerol included as a humectant, wash the tubing in running water for 3-4 hours. To remove any residual sulfur, treat the tubing with a 0.3% (w/v) solution of sodium sulfide at 80°C for 1 minute. Wash with hot water (60°C) for 2 minutes, followed by acidification with a 0.2% (v/v) solution of sulfuric acid, then rinse with hot water to remove the acid. Wetted membranes may be stored in water containing benzoic acid or formaldehyde.
D9777
Dialysis tubing cellulose membrane
avg. flat width 25 mm (1.0 in.)
Synonyme(s) :
Cellulose dialysis tubing, Dialysis membrane
Sélectionner une taille de conditionnement
289,00 $
Disponible pour expédition le09 avril 2025Détails
Sélectionner une taille de conditionnement
About This Item
289,00 $
Disponible pour expédition le09 avril 2025Détails
Produits recommandés
Diam. moyen
16 mm , when full
Largeur moy. à plat
25 mm (1.0 in.)
Capacité
~60 mL/ft
Vous recherchez des produits similaires ? Visite Guide de comparaison des produits
Catégories apparentées
Description générale
Conditionnement
Qualité
Notes préparatoires
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Vous ne trouvez pas la bonne version ?
Si vous avez besoin d'une version particulière, vous pouvez rechercher un certificat spécifique par le numéro de lot.
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Les clients ont également consulté
Contenu apparenté
Dialysis Tubing
-
Is this product ready to use?Do I need to boil it or place it into NaHCO3 before I use dialysis tube?
1 answer-
Helpful?
-
-
can it seprate 5 nm gold particle from gold solution
1 answer-
If the particle is larger than the tubing's 14 kDa molecular weight cut-off (MWCO) rating, it should be retained.
Helpful?
-
-
what is the storage temperature of this membrane
1 answer-
This product may be stored at room temperature. For products with no specified storage temperature given on the label or product page, ambient storage may be assumed. Products that require controlled temperature storage, such as refrigeration or freezer conditions, storage temperatures are published on the product label, Certificate of Analysis, and the Safety Data Sheet (SDS).
Helpful?
-
-
Hello, I would like to ask what is the pore size (in nanometres) of the dialysis membrane? (SKU code: D9777 ,Dialysis tubing cellulose membrane, avg. flat width 25 mm (1.0 in.))
1 answer-
The pore size is not determined for this product.
Helpful?
-
-
What are the molecular weight cut-off and the pores size of this dialysis tubing cellulose membrane?
1 answer-
As mentioned on the product detail page, the typical molecular weight cut-off is 14,000. However, this tubing will retain most proteins of molecular weight 12,000 or greater. The pore size has not been determined.
Helpful?
-
Active Filters
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique