Pular para o conteúdo
Merck
Todas as fotos(1)

Documentos Principais

G1007465

HPLC Analysis of Alfuzosin and Related Substances on Chromolith® HighResolution RP-18 endcapped

application for HPLC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

HPLC: suitable

parâmetros de teste

sample preparation: Dissolve 40.0 mg of the sample in 100 mL of mobile phase. 0.4 mg/mL (400 ppm) of USP Alfuzosin System Suitability Mixture RS in Mobile phase
column: Chromolith® HighResolution RP-18 endcapped, 150 x 4.6 mm (1.52023.0001)
mobile phase: [A] 5 ml of perchloric acid in 900 mL Milli-Q water. Adjust pH to 3.5 with 2M NaOH. Diluter to 1000 mL with water; [B] acetonitrile; [C] THF; (80:20:1, A:B:C)
flow rate: 1 mL/min
pressure: 994 psi (69 Bar)
column temp.: 25 °C
detector: 254 nm cell: 10 μL
injection: 10 μL
sample: 0.40 μg/mL of Alfuzosin Hydrochloride in Mobile phase from the Sample solution

adequação

application for HPLC

aplicação(ões)

pharmaceutical (small molecule)

Informações legais

Chromolith is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descrição

    Deacylated Alfuzosin
    Alfuzosin
    Furamide analog

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica