LC/MS/MS Analysis of Carbendazim in Orange Juice (Spiked) on Ascentis® Express C18 after Supel QuE PSA/C18 QuEChERS Cleanup

application for HPLC, application for SPE

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

HPLC: suitable
solid phase extraction (SPE): suitable

parâmetros de teste

sample preparation: Dispersive (QuEChERS)
sample/matrix: orange juice spiked with carbendazim at 1 ppb
extraction tube: Supel QuE Citrate (EN) Tube (55227-U)
extraction process: add 10 mL of orange juice with pulp to a 50 mL empty extraction tube (55248-U); add carbendazim at 1 ppb; add 10 mL acetonitrile; (shake for 1 minute; add contents of citrate extraction tube (55227-U); shake for 1 min; centrifuge at 3200 rpm for 5 minutes)
clean-up tube: Supel-Q PSA/C18 (55288-U)
clean-up process: transfer 0.7 mL of the acetonitrile layer to the PSA/C18 cleanup tube (55288-U); shake 1 minute; (centrifuge for 3 minutes at 3400 rpm; remove 0.25 mL of the supernatant; mix with 0.25 mL water for LC-MS analysis)
column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
mobile phase: [A] 10 mM ammonium acetate in water; [B] 10 mM ammonium acetate in methanol
gradient: 0-1 min: 30% B; 1.5-3.5 min: 100% B; 3.5-7 min: 30% B
flow rate: 0-1 min: 0.3 mL/min; 1.5-7 min: 0.5 mL/min
column temp.: 30 °C
detector: MS, ESI(+), MRM, m/z 192/160, 192/132
injection: 5 μL

adequação

application for HPLC
application for SPE

aplicação(ões)

food and beverages

Nota de análise

Simple and fast extraction and analytical method were developed for analysis of fungicide carbendazim in orange juice samples. QuEChERS extraction method was used with PSA/C18 cleanup. Ascentis Express C18 column chromatography was coupled to MS/MS detection with limits of detection at 0,1 ppb.

Informações legais

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

Analito

Descrição

    Carbendazim, PESTANAL®, analytical standard

Choose from one of the most recent versions:

Certificados de análise (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica