G003765
HPLC Analysis of Hydrocortisone in Topical Cream on Discovery® HS C18 after SPE using Discovery® DSC-Si
application for SPE, application for HPLC
About This Item
técnica(s)
HPLC: suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample addition: hydrocortisone topical hand cream in hexane:ethylacetate
SPE tube/cartridge: Discovery DSC-Si, 500 mg/3 mL (52695-U)
condition: 1 mL hexane:ethyl acetate (2:1, v/v)
sample addition: 1 mL
washing: 1 mL hexane:ethyl acetate (2:1, v/v)
elution: 1 mL methanol:DI water (1:1, v/v)
eluate post-treatment: dilute eluate to 2 mL with methanol:DI water (1:1, v/v)
column: Discovery HS C18, 15 cm x 4.6 mm I.D., 5 μm particles (568520-U)
mobile phase: [A] methanol; [B] DI water (1:1, A:B)
flow rate: 1 mL/min
column temp.: 35 °C
detector: UV, 220 nm
injection: 10 μL
adequação
application for HPLC
application for SPE
aplicação(ões)
cleaning products
cosmetics
personal care
pharmaceutical (small molecule)
Aplicação
Dissove 1 g 1% hydrocortisone topical hand cream in 10 mL ethyl acetate. Dilute 2 mL of ethyl acetate-cream sample with 8 mL hexane:ethyl acetate (2:1, v/v).
Informações legais
coluna analítica
tubo ou placa SPE
Choose from one of the most recent versions:
Certificados de análise (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica