Pular para o conteúdo
Merck
Todas as fotos(1)

Key Documents

CVSC03TP3

Millipore

Multimedia Durapore® 1.2/0.5/0.22 µm, Cartridge

cartridge nominal length 30 in. (75 cm), filter diam. 2.7 in. (6.9 cm), Code 0

Faça loginpara ver os preços organizacionais e de contrato


About This Item

Código UNSPSC:
23151806

Materiais

mixed cellulose esters (MCE) prefilter
polypropylene device
polypropylene support
silicone seal

Nível de qualidade

conformidade reg.

meets FDA 21 CFR 177-182 (all component materials)

esterilidade

non-sterile

Compatibilidade de esterilização

autoclavable compatible
steam-in-place compatible

linha de produto

Durapore®

Características

hydrophilic

fabricante/nome comercial

Durapore®

Parâmetros

≤32.4 mL/min air diffusion at 2.75 bar (40 psig) (in water)
0.35 bar max. differential pressure (5 psid) at 135 °C (forward)
0.35 bar max. inlet pressure (5 psi) at 135 °C (forward)
1.7 bar max. differential pressure (25 psid) at 80 °C (forward)
1.7 bar max. inlet pressure (25 psi) at 80 °C (forward)
3.5 bar max. differential pressure (50 psid) at 25 °C (reverse; intermittent)
3.5 bar max. inlet temp. (50 psi) at 25 °C (reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (forward)
5.5 bar max. inlet pressure (80 psi) at 25 °C (forward)

técnica(s)

sterile filtration: suitable

Comprimento

30 in.

Largura

2.7 in.

comprimento nominal do cartucho

30 in. (75 cm)

device size

30 in.

diâmetro do filtro

2.7 in. (6.9 cm)

filter filtration area

1.68 m2

Impurezas

≤0.5 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

extraíveis gravimétricos

≤70 mg/cartridge

matriz

Multimedia Durapore®

pore size

1.2/0.5/0.22 μm pore size

ponto de bolha

≥3450 mbar (50 psig), air with water

código

Code 0

Procurando produtos similares? Visita Guia de comparação de produtos

Descrição geral

Device configuration: Cartridge
Cartridge type: Liquid

Embalagem

Double Easy-Open bag

Nota de preparo

Sterilization Method: 6 autoclave or SIP cycles of 30 min at 123°C

This product was manufactured with a Durapore®  membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Nota de análise

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2  using ASTM® F838 methodology.
Gravimetric Extractables: The extractables level was ≤ 70 mg per 10-inch cartridge after 24 hours in water at controlled room temperature.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 15 L

Outras notas

Directions for Use:
  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer to Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Informações legais

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany

Não está encontrando o produto certo?  

Experimente o nosso Ferramenta de seleção de produtos.


Certificados de análise (COA)

Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica