- Development and validation of an ultra-performance liquid chromatography method for simultaneous analysis of 20 antihistaminics in dietary supplements.
Development and validation of an ultra-performance liquid chromatography method for simultaneous analysis of 20 antihistaminics in dietary supplements.
The purpose of this study was to develop and validate an ultra-performance liquid chromatography method for simultaneous analysis of 20 antihistamines (illegal additives) in dietary supplements. The limits of detection and quantitation of the method ranged from 1.5 to 2.5 µg/mL and from 20.0 to 50.0 µg/mL, respectively. The determination coefficient was >0.999, precisions were 0.2-5.1% (intra-day) and 0.1-8.8% (inter-day), and accuracies were 84.5-111.2% (intra-day) and 91.9-112.0% (inter-day). The mean recoveries of 20 targeted compounds from dietary supplements ranged from 75.4 to 119.3%. The relative standard deviations were <6.6% and complied with established international guidelines. The relative standard deviation of stability was <0.8%. Fifty-two commercially available dietary supplements were evaluated using this method, and were found to have none of the 20 antihistamines in significant abundance.