Methode(n)

HPLC: suitable

Testparameter

column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] water; [B] acetonitrile
gradient: 20 to 100% B in 5 min; held at 100% B for 0.5 min
flow rate: 0.3 mL/min
pressure: 2750 psi (190 bar)
column temp.: 35 °C
detector: UV, 227 nm
injection: 5 μL
sample: 20 mg/mL in water:acetonitrile (50:50); sonicate for 15 minutes; filter 0.45 μm; dilute to water:acetonitrile (80:20)

Eignung

application for HPLC

Anwendung(en)

vitamins, nutraceuticals, and natural products

Hinweis zur Analyse

This application demonstrates the suitability of Ascentis Express F5 for the efficient separation of withania extract.

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyt

Beschreibung

    Withanoside IV
    Withanoside V
    Withaferin A
    12-Deoxywithastramonolide
    Withanolide A
    Withanone
    Withanolide B

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Choose from one of the most recent versions:

Analysenzertifikate (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Kundensupport

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.