HPLC Analysis of Vitamin D2 and D3 on Ascentis® Express C18
![HPLC Analysis of Vitamin D2 and D3 on Ascentis® Express C18 application for HPLC](/deepweb/assets/sigmaaldrich/product/images/356/809/4184914a-f7ee-4baa-8523-aba285b1118c/800/4184914a-f7ee-4baa-8523-aba285b1118c.jpg)
Materiales
analytical column
Referencia del producto
Descripción
Precios
CONDITIONS
column
Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)
mobile phase
acetonitrile
flow rate
0.15 mL/min
column temp.
35 °C
detector
UV, 290 nm
injection
10 μL
sample
10 mg/L each (vitamins D2 and D3) in methanol