US EPA Method OLM04.2 SVOA: GC Analysis of Semivolatiles on SLB®-5ms
Materiales
analytical column
For use with
Referencia del producto
Descripción
Precios
Wheaton wide-mouth bottle
amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 eaguard column
Referencia del producto
Descripción
Precios
standard
Referencia del producto
Descripción
Precios
EPA 8270 GC-MS Tuning Solution II
certified reference material, 1000 μg/mL each component in dichloromethaneCLP Semivolatile Calibration Mix
certified reference material, 1000 μg/mL each component in dichloromethane : benzene (3:1)CLP Semivolatiles Surrogate Standards Mix
certified reference material, 4000 μg/mL each component in methylene chlorideSemivolatile Internal Standards Mix
certified reference material, TraceCERT®, 2000 μg/mL each component in methylene chloride, ampule of 2 × 1 mLCONDITIONS
column
SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
40 °C (2 min), 15 °C/min to 240 °C, 10 °C/min to 330 °C (2 min)
inj. temp.
250 °C
MSD interface
330 °C
scan range
m/z = 40-450
carrier gas
helium, 1.0 mL/min (11 min), ramp to 1.5 mL/min (constant for remainder of run)
injection
1 μL, pulsed splitless (15 psi until 0.10 min), (0.75 min)
liner
4 mm I.D., single taper
sample
50 ppm of an 80 component semivolatile standard, plus 6 internal standards (at 40 ppm) in methylene chloride
Descripción
Legal Information
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany