GC Analysis of Class 2 Residual Solvents on Equity-1 after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Class 2 Residual Solvents on Equity-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC

Materiales

SPME fiber

Referencia del producto
Descripción
Precios

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 11) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

85 μm Polyacrylate, 24 ga (57304)

extraction

eated headspace, 60 °C for 5 min

desorption flow

250 °C for 3 min

column

Equity-1, 10 m x 0.20 mm I.D., 1.2 μm (28043-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °

injection

5:1 split

liner

0.75 mm I.D. SPME type