G006151
UHPLC/MS Analysis of Antineoplastic and Antidepressant Drugs in Plasma on Titan C18 after SPE using HybridSPE®-PLus
application for HPLC, application for SPE
About This Item
Productos recomendados
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample/matrix: rat plasma
SPE well plate: HybridSPE-PLus 96-Well Plate, 50 mg/2 mL per well (575659-U)
sample preparation: final analyte concentration 50 ng/mL
sample addition: 400 μL of 1% formic acid acetonitile:spiked plasma (3:1) per well
column: Titan C18, 5 cm x 2.1 mm I.D., 1.9 μm particles (577122-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in 90:10 (v:v) acetonitrile:water
gradient: 35 to 90% B in 1.5 min, held at 90% B for 0.4 min
flow rate: 0.4 mL/min
pressure: 3582 psi (247 bar)
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
suitability
application for HPLC
application for SPE
application(s)
pharmaceutical (small molecule)
Analysis Note
Legal Information
Analito
SPE tube or plate
standard
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico