技術
HPLC: suitable
測試參數
column: BIOshell A160 Peptide C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (66905-U)
mobile phase: [A] 70:30, water with 0.1% TFA:acetonitrile with 0.09% TFA; [B] 55:45, water with 0.1% TFA:acetonitrile with 0.09% TFA
gradient: 0 to 100% B in 7.5 min
flow rate: 0.3 mL/min
pressure: 4040 psi (278 bar)
column temp.: 30 °C
detector: UV, 215 nm
injection: 0.5 μL
sample: 200 μg/mL insulin (bovine, human, porcine) in 0.1% formic acid
適合性
application for HPLC
應用
life science and biopharma
pharmaceutical (small molecule)
分析報告
These three insulins have more than 76% identical positions in a primary structure alignment. The human and porcine proteins are most similar with 85% identity. Baseline resolution of these insulins in achieved on BIOshell A160 Peptide C18, with relative band spacing of these three peaks approximating their relative sequence similarities. BIOshell A160 Peptide C18 is ideally suited for RP chromatography of small proteins or large peptides.
法律資訊
分析物
描述
Insulin from bovine pancreas, ≥25 USP units/mg (HPLC), powder
Insulin human, powder, recombinant, expressed in E. coli, suitable for cell culture
Insulin from porcine pancreas, powder, ≥27 USP units/mg
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務