HPLC Analysis of Cosmetic Preservatives on Ascentis® Express RP-Amide

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis Express RP-Amide, 10 cm x 3 mm I.D., 2.7 μm particles (53918-U)
mobile phase: [A] water; [B] methanol; (60:40, A:B)
flow rate: 0.6 mL/min
pressure: 2900 psi (200 bar)
column temp.: 50 °C
detector: UV, 250 nm
injection: 1 μL
sample: methyl-, benzylparaben (50 mg/L), ethylparaben (30 mg/L), propyl-, butylparaben (100 mg/L), 2-phenoxyethanol (1.5 g/L),1-phenoxy-2-propanol (3 g/L), all in 40% methanol

適合性

application for HPLC

應用

cleaning products
cosmetics
personal care

分析報告

With the publication Borremans, M. et.al. 2004, Chromatographia 59: 47-53 as a reference, this method is 10% faster, but with superior resolution of critical pairs; in this case, the minimum USP resolution is 2.4.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

描述

    Ethylene glycol monophenyl ether
    1-Phenoxy-2-propanol
    Methyl 4-hydroxybenzoate
    Ethyl 4-hydroxybenzoate
    Butylparaben
    Benzyl 4-hydroxybenzoate