技術
gas chromatography (GC): suitable
測試參數
column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 60 °C (1 min), 10 °C/min to 330 °C (10 min)
inj. temp.: 250 °C
carrier gas: helium, 0.6 mL/min, constant
liner: 4 mm ID focus liner with taper
sample: cologne sample spiked at 100 μg/mL with various phthalate esters
應用
cleaning products
cosmetics
personal care
一般說明
Diethyl phthalate was detected in the unspiked sample.
分析報告
Phthalate esters are commonly used as plasticizers and additives for various materials. In the case of some fragrances, diethyl phthalate is used as a solvent and deluent. Rising concern with the potential health effects of exposure to phthalate esters has lead to some controversy over their use in fragrances and other personal care products. This application illustrates the use of GC-MS to analyze a cologne sample that has been spiked at 100 ug/ml with a list of phthalate esters. The unspiked sample was found to contain diethyl phthalate.
The SLB-5ms column was chosen for the analysis due to its low bleed/MS compatibility. The nonpolar character of the phase resulted in elution of the lower boiling constituents in the cologne sample prior to most of the phthalate esters. A 20 m x 0.18 mm ID x 0.18 μm column was used to reduce overall analysis time.
The SLB-5ms column was chosen for the analysis due to its low bleed/MS compatibility. The nonpolar character of the phase resulted in elution of the lower boiling constituents in the cologne sample prior to most of the phthalate esters. A 20 m x 0.18 mm ID x 0.18 μm column was used to reduce overall analysis time.
法律資訊
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務