推薦產品
技術
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
測試參數
sample/matrix: 3 g of 2% milk + 10 μL internal standard solution (20 μg/mL 4-methyl-2-pentanone) (9 mL GC vial)
SPME fiber: PDMS/Carboxen, 75 μm (57318)
extraction: headspace, 15 min with constant stirring at 45 °C
desorption process: 5 min, 250 °C
column: Supel-Q PLOT, 30 m × 0.32 mm I.D. (24242)
oven: 70 °C (2 min) to 140 °C at 6 °C/min (2 min hold) then to 220 °C at 6 °C/min (5 min hold)
inj. temp.: 250 °C
detector: GC-MS ion trap, m/z = 33-300
carrier gas: helium, 35 cm/sec
injection: splitless, closed 2 min
liner: 0.75 mm liner
sample: Milk
適合性
application for SPME
應用
food and beverages
分析報告
In this study, SPME was used for the analysis of volatiles in milk using a headspace method. Light can catalyze the formation of peroxides that can cleave double bonds in fatty acids. The cleaved by-products form aldehydes that can give the food a pungent off flavor. Dr. Ray Marsili investigated the feasibility of using SPME as a substitute sample preparation device prior to GC/MS analysis. The Carboxen/PDMS fiber was evaluated for the extraction of low ppb levels of pentanal, hexanal, and dimethyldisulfide. The results of the comparison of SPME to dynamic headspace analysis indicate that SPME-GC/MS is more accurate and precise while providing sensitivities equivalent to DH-GC/MS. The GC column used, Supel-Q PLOT, operates via gas-solid chromatography with a DVB adsorbent and permits the separation of a wide range of analytes.
法律資訊
分析物
描述
4-Methyl-2-pentanone
Dimethyl disulfide
Hexanal
3-Methylpentane
Acetone
固相微萃取纖維
標準
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務