推薦產品
參數
0-100 psi pressure range
技術
gas chromatography (GC): suitable
一般說明
Install a pressure gauge in the gas delivery system to measure pressure at any given point. For example, install one upstream and downstream of the High Capacity Gas Purifier. When the pressure drop across the purifier exceeds 10 psi, the converter tube should be replaced. Kit contains:
- 2 in. (5 cm) diameter steel and copper alloy gauge (0-100 psi)
- NPT to Swagelok adapter
- 1/8 in. tee
- 18 in. (1/2 m) of 1/8 in. copper line
- assembly and installation instructions
應用
In-line pressure gauge kit may be used to monitor the column inlet pressure of high-performance liquid chromatography (HPLC) equipment, which is further modified for carbon-dioxide supercritical fluid chromatography (SFC).
客戶也查看了
Modification of conventional high-performance liquid chromatography equipment for use in carbon dioxide supercritical fluid chromatography.
Journal of Chromatography A, 371, 109-119 (1986)
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務