SLFGR25LS
Millex® hydrophobic PTFE syringe filter
pore size 0.2 μm, diam. 25 mm, sterile
Synonym(s):
Millex®-FG 0.2 µm Filter Unit, Millex®-FG Filter Unit or Vent Filter, vent filter, air filter, gas filter
About This Item
Recommended Products
material
PTFE membrane
PVC housing
Quality Level
sterility
ethylene oxide treated
sterile
product line
Millex®
feature
hydrophobic
packaging
pack of 50 ea
manufacturer/tradename
Millex®
technique(s)
gas filtration: suitable
sterile filtration: suitable
vent filtration: suitable
diam.
25 mm
filtration area
4 cm2
housing diam.
29 mm
inlet to outlet H
25 mm
volume
100 mL
matrix
Fluoropore®
pore size
0.2 μm
input
mild organic solvent(s)
gas
bubble point
≥13 psi, methanol
fitting
female Luer-Lok® inlet
male Luer outlet slip
General description
Specificity
Membrane: Hydrophobic Fluoropore® polytetrafluoroethylene (PTFE), type FG
Pore size: 0.2 μm
Housing: Polyvinyl chloride (PVC)
Dimensions
Inlet to outlet: 25 mm (0.98 in.)
Diameter: 29 mm (1.14 in.)
Filtration area: 4 cm2 (0.62 in2)
Temperature Limit: 45 °C (113 °F)
Pressure limit at 21 °C: 5.2 bar (75 psi) inlet and differential maximum
Air flow rate: ≥ 1.0 L/min at 1 psi
Sterilization method: Ethylene oxide gas
Connections Female Luer-Lok® inlet; male Luer-slip outlet
Methanol bubble point: ≥ 13 psi
Color: blue inlet, blue outlet
Application
Features and Benefits
- For research use only (RUO)
- Single use only
- Broad chemical compatibility
- Non-pyrogenic and non-toxic
Legal Information
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service