跳转至内容
Merck

LC/MS/MS Analysis of Tetracycline Antibiotics on Kromasil® Eternity C18

LC/MS/MS Analysis of Tetracycline Antibiotics on Kromasil® Eternity™ C18 application for LC-MS

材料

分析柱

产品编号
说明
价格

Kromasil® Eternity C18 高效液相色谱柱

2.5 μm particle size, pore size 100 Å, L × I.D. 100 mm × 2.1 mm

標準

产品编号
说明
价格

金霉素盐酸盐 盐酸盐

VETRANAL®, analytical standard

土霉素盐酸盐 盐酸盐

VETRANAL®, analytical standard

4-差向四环素 盐酸盐

VETRANAL®, analytical standard

多西环素 单盐酸半乙醇半水合物

VETRANAL®, analytical standard

四环素 盐酸盐

VETRANAL®, analytical standard

CONDITIONS

column

Kromasil Eternity C18, 10 cm x 2.1 mm I.D., 2.5 μm particles (K08670895)

mobile phase

[A] 0.5% acetonitrile and 0.1% formic acid in water; [B] 0.1% formic acid in methanol:acetonitrile (1:1)

gradient

Held at 10% B for 0.1 min; to 20% B in 1.9 min; held at 20% B for 3.5 min; to 80% B in 0.1 min; held at 80% B for 1.0 min; to 10% B in 0.1 min; held at 10% B for 3.3 min

flow rate

0.2 mL/min

pressure

3070 psi (212 bar)

column temp.

30 °C

detector

MS/MS

injection

10 μL

sample

each compound 50 ng/mL in acetonitrile:water (11:89)

说明

分析報告

The analysis of tetracycline antibiotics can be problematic via liquid chromatography-mass spectrometry due to their chelating properties resulting in compounds′ irreversible retention. This application uses the organosilane-coated silica stationary phase that prevents unwanted binding of compounds and provides good peak shape with an LC/MS-compatible mobile phase. The method also delivers good resolution of compounds from their isobaric epi-metabolites that cannot be resolved via mass spectrometry alone.