跳转至内容
Merck

GC Analysis of Aldehydes (PFBOA Derivatives) in Beer on SLB®-5ms after SPME using a 65 μm PDMS/DVB Fiber

GC Analysis of Aldehydes (PFBOA Derivatives) in Beer on SLB®-5ms after SPME using a 65 μm PDMS/DVB Fiber suitable for GC, application for SPME

材料

分析柱

产品编号
说明
价格

SLB ®-5 ms 毛细管 GC 色谱柱

L × I.D. 30 m × 0.25 mm, df 0.50 μm

固相微萃取纖維

产品编号
说明
价格

SPME固相微萃取萃取头,二甲基硅氧烷/二乙烯基苯(PDMS/DVB)

df 65 μm(PDMS/DVB, for use with autosampler, needle size 23 ga

標準

产品编号
说明
价格

苯甲醛

analytical standard

苯乙醛

≥95%, FCC, FG

异丁醛

analytical standard

2-甲基丁醛

analytical standard

异戊醛

analytical standard

反式-2-壬烯醛

analytical standard

顺-11-十六碳醛

analytical standard

糠醛

certified reference material, pharmaceutical secondary standard

相關產品

产品编号
说明
价格

SPME 光纤支架

for use with manual sampling

SPME 光纤支架

for use with CTC CombiPAL and Gerstel MPS2

CONDITIONS

sample preparation

derivatization procedure: Add 100 μL of PFBOA solution (6 g/L) to 10 mL of deionized water in a 20 mL glass vial and seal. (Insert the PDMS/DVB SPME fiber in the headspace of the PFBOA solution for 10 min at 50 °C.)

sample/matrix

10 mL of degassed beer, 3.5 gm sodium chloride in 20 mL vial

SPME fiber

PDMS-DVB, 23 guage, 65 μm, Auto, Pk/3 (57345-U)

extraction

60 min, headspace, 50 °C, agitation (5 min on, 30 min off)

desorption process

0.10 min, 250 °C

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.50 μm (28473-U)

oven

40 °C, 7 °C/min to 250 °C (14 min)

inj. temp.

250 °C

detector

MSD, SIM

MSD interface

275 °C

carrier gas

helium, 1.1 mL/min

injection

splitless

liner

0.75 mm I.D. SPME liner

sample

Beer

说明

分析報告

The formation of aldehydes is a major contributor to the deterioration in flavor or "staling" of beer upon storage. Research indicates that the source of these aldehydes is the type of malt used and quality of the wort produced during the brewing process. This application illustrates the use of SPME to extract aldehydes from beer, followed by on-fiber derivatization with 0-(2,3,4,5,6- pentafluorobenzyl)hydroxylamine hydrochloride (PFBOA). The aldehydes are then analyzed by GC/MS-SIM as their corresponding PFBOA derivatives. The aldehydes exist as geometric isomers, resulting in two derivatives for each. In the case of several of the analytes, these two derivatives were resolved chromatographically.

法律資訊

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany