跳转至内容
Merck

US EPA Method 8081: GC Analysis of Organochlorine Pesticides on Equity-5

US EPA Method 8081: GC Analysis of Organochlorine Pesticides on Equity-5 suitable for GC

材料

分析柱

产品编号
说明
价格

Equity-5 毛细管GC色谱柱

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

Equity-5, 30 m × 0.25 mm I.D., 0.25 μm (28089-U)

oven

100 °C (2 min), 15 °C/min to 160 °C, 5 °C/min to 300 °C (10 min)

inj. temp.

225 °C

detector

ECD, 310 °C

carrier gas

helium, 30 cm/sec @ 100 °C

injection

2 μL, splitless (0.5 min)

liner

4 mm I.D., double taper

sample

50 ppb of a 22-component chlorinated pesticides standard (46845-U)

说明

一般說明

As per United States Environmental Protection Agency (EPA), method 8081 may be used in determining the concentrations of organochlorine pesticides in solid and liquid extracts. It can be achieved by using fused-silica, open tubular, capillary columns. This method allows choosing between single column and a dual column.