HPLC Analysis of Vitamin D2 and D3 on Ascentis® Express C18

application for HPLC

登录查看公司和协议定价


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis® Express C18, 25 cm x 4.6 mm I.D., 5 μm particles (50538-U)
mobile phase: [A] acetonitrile; [B] methanol; (95:5, A:B)
flow rate: 1.5 mL/min
pressure: 1726 psi (119 bar)
column temp.: 35 °C
detector: UV, 265 nm
injection: 10 μL
sample: 50 μg/mL in 90:10, acetonitrile:methanol

適合性

application for HPLC

應用

clinical
vitamins, nutraceuticals, and natural products

分析報告

This application demonstrates the suitability of the Ascentis Express C18, 5 μm, for the routine analysis of the fat-soluble vitamins D2 and D3 . Structures along with the optimized chromatogram obtained on the Ascentis Express C18, 5 μm are presented.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Ergocalciferol, 40,000,000 USP units/g
    Cholecalciferol, ≥98% (HPLC)

Choose from one of the most recent versions:

分析证书(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客户支持