推荐产品
技術
gas chromatography (GC): suitable
測試參數
column: SUPELCOWAX 10, 15 m x 0.10 mm I.D., 0.10 μm (24343)
oven: 80°C (0.3 min), 30 °C/min to 180 °C (0.5 min), 10 °C/min to 200 °C (1 min)
inj. temp.: 250 °C
detector: FID, 250 °C
carrier gas: hydrogen, 45 cm/sec, constant
injection: 0.1 μL 200:1 split
liner: 2 mm I.D., straight
sample: bacterial methyl ester standards in methyl caproate, 10 mg/mL (47080-U)
應用
clinical
分析報告
Food-borne microbial pathogens can cause illness in humans. Because each bacterium has a unique cellular fatty acid profile, GC analysis can be used to help with identification. Analysis is performed following derivatization of cellular fatty acids to methyl esters. These are called bacterial acid methyl esters (BAMEs) instead of fatty acid methyl esters (FAMEs) to signify the bacterial source. The chromatogram shown is the analysis of a standard mix containing 26 BAME compounds. The polar nature of the column, along with the use of Fast GC column dimensions and instrument conditions, allows good peak shape and resolution in just 7 minutes.
法律資訊
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门