The S-Traps are not included with any of the Oxygen/Moisture traps and have been discontinued. These traps have a 1/8 inch Swagelok fitting.
Przejdź do
Wybierz wielkość
| Rozmiar/SKU | Dostępność | Cena netto |
|---|
Informacje o tej pozycji
technique(s)
gas chromatography (GC): suitable using :Suitable
volume
120 cc
fitting
1/8 in. fitting
General description
Dimensions:
- 120 cc Trap: 11
1/8 in. (28.2 cm) long x 13/8 in. (35 mm) diameter - 750 cc Trap: 16
1/2 in. (41.9 cm) long x 25/16 in. (59 mm) diameter - S-Trap: 7
1/2 in. (19 cm) long x 59/16 in. (14.1 cm) wide x1/2 in. (13 mm) diameter [total bed length = 193/4 in. (50.2 cm); the extended bed length ensures prolonged contact between the gas and the adsorbent and provides greater working capacity]
Application
Legal Information
1 of 1
Ta pozycja | |||
|---|---|---|---|
| description volume 120 cc, 1/8 in. fitting | description volume 750 cc, 1/4 in. fitting | description volume 750 cc, 1/2 in. fitting | description volume 120 cc, 1/4 in. fitting |
| volume 120 cc | volume 750 cc | volume 750 cc | volume 120 cc |
| technique(s) gas chromatography (GC): suitable using :Suitable | technique(s) gas chromatography (GC): suitable using :Suitable | technique(s) - | technique(s) gas chromatography (GC): suitable using :Suitable |
| fitting 1/8 in. fitting | fitting 1/4 in. fitting | fitting 1/2 in. fitting | fitting 1/4 in. fitting |
Still not finding the right product?
Explore all of our products under Supelpure-O Oxygen/Moisture Trap
Słowo ostrzegawcze
Danger
Kody zagrożeń
Hazard Classifications
Aquatic Chronic 2 - Carc. 1A Inhalation - Resp. Sens. 1 - Skin Sens. 1 - STOT RE 2 Inhalation
Narządy docelowe
Lungs
Klasa składowania
6.1D - Non-combustible acute toxic Cat.3 / toxic hazardous materials or hazardous materials causing chronic effects
Temperatura zapłonu (°F)
Not applicable
Temperatura zapłonu (°C)
Not applicable
Wykazy regulacyjne
Wykazy regulacyjne dotyczą głównie produktów chemicznych. Można w nich podawać ograniczoną liczbę informacji na temat produktów niechemicznych. Brak wpisu oznacza, że żaden ze składników nie znajduje się w wykazie. Użytkownik odpowiada za zagwarantowanie bezpiecznego i zgodnego z prawem stosowania produktu.
1313-99-1
Wybierz jedną z najnowszych wersji:
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
-
Is the S-Trap included with purchase of either the 750cc or 120cc trap? What style of fitting do the traps have?
1 answer-
Helpful?
-
-
What is the expected lifetime of this mositure trap?
1 answer-
Product No. 22449 has a stated life of 6.5 gas cylinders. This stated life is based on 9 in. x 51 in. cylinders (218 cubic feet) of carrier gas containing 50 ppm of contaminants supplying one chromatograph with two capillary columns. To correct for other volumes of gas per cylinder, multiply this stated life (cylinders) by the number of cubic feet in the cylinders, then divide by 218.
Helpful?
-
-
Regarding the Supelpure™-O Oxygen/Moisture Trap 120 cc, what flow rate is suitable for this device? And is removal efficiency dependent on temperature? At last, could I use this device to remove ppm level oxygen contaminations in pure CO2 as well?
1 answer-
Please see below for responses to each question submitted.
1. What flow rate is suitable for this device?
The maximum flow rate for the Supelpure™-O Oxygen/Moisture Trap - 22449 is 250 ml/minute. The pressure setting on the incoming carrier gas tank depends on what type of analysis is being done (packed or capillary, isothermal or temperature programmed method, etc.) The typical recommendation for GC applications is that the pressure be at least 20 psi higher than that achieved during the analysis - preferably 40-50 psi higher.2. Is removal efficiency dependent upon temperature?
Unless the GC maintains constant pressure, as the temperature goes up, resistance/increased back pressure causes the flow rate to drop.3. Is this device able to remove ppm level oxygen contaminations in pure CO2
The capacity for these traps will not vary considerably with changes in the inlet concentration of O2. At high levels of O2, the efficiency may be impacted. Oxygen is reduced to less than 0.5 ppm when the level of the incoming gas does not exceed 10 ppm. Since this trap can only remove oxygen and moisture, it can be used to remove oxygen from the CO2 gas stream as well.Helpful?
-