HPLC Analysis of Triacylglycerols in Olive Oil on Discovery® HS C18 after Sample Prep using Discovery® DSC-Si (Propionitrile Mobile Phase, ELSD Detection)
Materiales
analytical column
Referencia del producto
Descripción
Precios
SPE tube or plate
Referencia del producto
Descripción
Precios
mobile phase component
Referencia del producto
Descripción
Precios
CONDITIONS
SPE tube/cartridge
Discovery DSC-Si, 1 gm/6 mL (52656-U)
condition
6 mL hexane
sample addition
0.12 g oil in 0.5 mL hexane
elution
10 mL hexane-diethyl ether 87-13 (v/v)
eluate post-treatment
evaporation to dryness, reconstituted into 2 mL acetone
column
Discovery HS C18, 25 cm x 4.6 mm I.D., 5 μm particles (568523-U)
mobile phase
propionitrile
flow rate
1.5 mL/min
pressure
1117 psi (77 bar)
column temp.
30 °C
detector
ELSD
injection
5 μL
Descripción
Analysis Note
Olive oil was pre-cleaned using Silica SPE, analyzed as 5% sample in acetone (w/v).
Legal Information
Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany