跳转至内容
Merck

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

分析柱

产品编号
说明
价格

Equity-1 毛细管GC色谱柱

L × I.D. 30 m × 0.32 mm, df 1.00 μm

固相微萃取纖維

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial

SPME fiber

PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)

extraction

headspace, 60 °C for 5 min

desorption process

250 °C for 0.75 min

column

Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)

oven

50 °C

inj. temp.

250 °C

detector

FID, 200 °C

carrier gas

helium, 30 cm/sec

liner

0.75 mm I.D., SPME type, straight design (unpacked)

说明

法律資訊