추천 제품
기술
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
검사 파라미터
sample/matrix: corn oil spiked with phthalate esters at 500 μg/kg
SPME fiber: 100 μm PDMS (57300-U)
extraction: headspace, 95 °C (measured in headspace)
desorption process: 260 °C for 4 min
column: F.S. CAP. SLB-5ms 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 60 °C (1 min), 10 °C/min to 330 °C (10 min)
inj. temp.: 260 °C
carrier gas: helium, 0.6 mL/min, constant
liner: 0.75 mm I.D. SPME
sample: Corn oil
적합성
application for SPME
응용 분야
agriculture
food and beverages
분석 메모
In this application, SPME was used for the analysis of phthalate esters from oily samples, specifically the oil-based flavor packets included with ramen noodle kits. Spiked samples of corn oil and flavored oils included with ramen noodle kits were analyzed using high temperature SPME headspace extraction using a 100 μm PDMS fiber which was then directly desorbed into the GC/MS system. Calibration standards were prepared in corn oil, and spiked samples of chicken and beef oil-based flavors were analyzed. The SLB-5ms GC columns provide consistently low-bleed, inert, efficient, and durable separations.