GC Analysis of Phthalate Esters in Corn Oil (Spiked) on SLB®-5ms after SPME with a 100 μm PDMS Fiber

suitable for GC, application for SPME

로그인조직 및 계약 가격 보기


About This Item

기술

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

검사 파라미터

sample/matrix: corn oil spiked with phthalate esters at 500 μg/kg
SPME fiber: 100 μm PDMS (57300-U)
extraction: headspace, 95 °C (measured in headspace)
desorption process: 260 °C for 4 min
column: F.S. CAP. SLB-5ms 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 60 °C (1 min), 10 °C/min to 330 °C (10 min)
inj. temp.: 260 °C
carrier gas: helium, 0.6 mL/min, constant
liner: 0.75 mm I.D. SPME
sample: Corn oil

적합성

application for SPME

응용 분야

agriculture
food and beverages

분석 메모

In this application, SPME was used for the analysis of phthalate esters from oily samples, specifically the oil-based flavor packets included with ramen noodle kits. Spiked samples of corn oil and flavored oils included with ramen noodle kits were analyzed using high temperature SPME headspace extraction using a 100 μm PDMS fiber which was then directly desorbed into the GC/MS system. Calibration standards were prepared in corn oil, and spiked samples of chicken and beef oil-based flavors were analyzed. The SLB-5ms GC columns provide consistently low-bleed, inert, efficient, and durable separations.

법적 정보

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

분석 컬럼


가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.