The S-Traps are not included with any of the Oxygen/Moisture traps and have been discontinued. These traps have a 1/8 inch Swagelok fitting.
추천 제품
용적
120 cc
피팅
1/8 in. fitting
유사한 제품을 찾으십니까? 방문 제품 비교 안내
일반 설명
Dimensions:
- 120 cc Trap: 111/8in. (28.2 cm) long x 13/8 in. (35 mm) diameter
- 750 cc Trap: 161/2 in. (41.9 cm) long x 25/16in. (59 mm) diameter
- S-Trap: 71/2 in. (19 cm) long x 59/16 in. (14.1 cm) wide x 1/2 in. (13 mm) diameter [total bed length = 193/4 in. (50.2 cm); the extended bed length ensures prolonged contact between the gas and the adsorbent and provides greater working capacity]
애플리케이션
법적 정보
신호어
Danger
유해 및 위험 성명서
Hazard Classifications
Aquatic Acute 1 - Aquatic Chronic 1 - Carc. 1A Inhalation - Resp. Sens. 1 - Skin Sens. 1 - STOT RE 2 Inhalation
표적 기관
Lungs
Storage Class Code
6.1D - Non-combustible acute toxic Cat.3 / toxic hazardous materials or hazardous materials causing chronic effects
WGK
WGK 3
Flash Point (°F)
Not applicable
Flash Point (°C)
Not applicable
가장 최신 버전 중 하나를 선택하세요:
이미 열람한 고객
-
Is the S-Trap included with purchase of either the 750cc or 120cc trap? What style of fitting do the traps have?
1 답변-
도움이 되었습니까?
-
-
What is the expected lifetime of this mositure trap?
1 답변-
Product No. 22449 has a stated life of 6.5 gas cylinders. This stated life is based on 9 in. x 51 in. cylinders (218 cubic feet) of carrier gas containing 50 ppm of contaminants supplying one chromatograph with two capillary columns. To correct for other volumes of gas per cylinder, multiply this stated life (cylinders) by the number of cubic feet in the cylinders, then divide by 218.
도움이 되었습니까?
-
-
Regarding the Supelpure™-O Oxygen/Moisture Trap 120 cc, what flow rate is suitable for this device? And is removal efficiency dependent on temperature? At last, could I use this device to remove ppm level oxygen contaminations in pure CO2 as well?
1 답변-
Please see below for responses to each question submitted.
1. What flow rate is suitable for this device?
The maximum flow rate for the Supelpure™-O Oxygen/Moisture Trap - 22449 is 250 ml/minute. The pressure setting on the incoming carrier gas tank depends on what type of analysis is being done (packed or capillary, isothermal or temperature programmed method, etc.) The typical recommendation for GC applications is that the pressure be at least 20 psi higher than that achieved during the analysis - preferably 40-50 psi higher.2. Is removal efficiency dependent upon temperature?
Unless the GC maintains constant pressure, as the temperature goes up, resistance/increased back pressure causes the flow rate to drop.3. Is this device able to remove ppm level oxygen contaminations in pure CO2
The capacity for these traps will not vary considerably with changes in the inlet concentration of O2. At high levels of O2, the efficiency may be impacted. Oxygen is reduced to less than 0.5 ppm when the level of the incoming gas does not exceed 10 ppm. Since this trap can only remove oxygen and moisture, it can be used to remove oxygen from the CO2 gas stream as well.도움이 되었습니까?
-