The capture antibody will detect SAA1 and SAA2. The detection will also detect SAA1 but does not detect SAA2. Neither antibody has been tested against SAA3 or SAA4.
HCVD2MAG-67K
MILLIPLEX® Human Cardiovascular Disease (CVD) Magnetic Bead Panel 2 - Cardiovascular Disease Multiplex Assay
The analytes available for this multiplex kit are: ADAMTS13, D-Dimer, GDF-15, Myoglobin, sICAM-1, MPO, P-selectin, Lipocalin-2/NGAL, sVCAM-1, SAA.
About This Item
추천 제품
Quality Level
종 반응성
human
제조업체/상표
Milliplex®
assay range
accuracy: 88%
(Myoglobin)
accuracy: 90%
(P-selectin)
accuracy: 98%
(GDF-15)
standard curve range: 0.0005-2.5 ng/mL
(GDF-15)
standard curve range: 0.012-50 ng/mL
(Lipocalcin-2/NGAL)
standard curve range: 0.024-100 ng/mL
(Myoglobin & MPO)
standard curve range: 0.085-350 ng/mL
(sICAM-1)
standard curve range: 0.122-500 ng/mL
(sVCAM-1)
standard curve range: 0.244-1,000 ng/mL
(D-dimer, SAA & P-selectin)
standard curve range: 1.221-5,000 ng/mL
(ADAMTS 13)
inter-assay cv: <20%
intra-assay cv: <15%
기술
multiplexing: suitable
검출 방법
fluorometric (Luminex xMAP)
배송 상태
wet ice
일반 설명
MILLIPLEX® Human Cardiovascular Disease (CVD) Panel 2 is a 10-plex kit to be used for the simultaneous quantification of any or all of the following analytes in human serum, plasma and tissue/cell lysate and culture supernatant samples: Adamts13, D-dimer, GDF-15, Myoglobin, sICAM-1, MPO, P-selectin, Lipocalin-2/NGAL, sVCAM-1 and SAA. This kit uses a 96-well format, contains a lyophilized standard cocktail, two internal assay quality controls and can measure up to 38 samples in duplicate.
The Luminex® xMAP® platform uses a magnetic bead immunoassay format for ideal speed and sensitivity to quantitate multiple analytes simultaneously, dramatically improving productivity while conserving valuable sample volume.
Panel Type: Cardiovascular
특이성
There was no or negligible cross-reactivity between the antibodies for an analyte and any of the other analytes in this panel.
애플리케이션
- Analytes: ADAMTS13, D-Dimer, GDF15, sICAM-1, Myeloperoxidase (MPO), Myoglobin, NGAL/Lipocalin-2, sP-selectin, Serum Amyloid A (SAA), sVCAM-1
- Recommended Sample Type: Human serum, plasma, cell/tissue culture supernatants and lysates
- Recommended Sample Dilution: 25 μL per well of 1:100 diluted serum or plasma; tissue/cell culture samples may require dilution in appropriate control medium
- NOTE: D-Dimer cannot be measured in plasma samples
- Assay Run Time: Overnight (16-18 hours) at 2-8°C or 2 hours at room temperature (20-25 °C)
- Research Category: Cardiovascular Disease
- Research Subcategory: Metabolic Disorders
특징 및 장점
기타 정보
법적 정보
면책조항
신호어
Warning
유해 및 위험 성명서
Hazard Classifications
Acute Tox. 4 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Aquatic Chronic 2 - Eye Irrit. 2 - Skin Irrit. 2 - Skin Sens. 1 - STOT SE 3
표적 기관
Respiratory system
Storage Class Code
10 - Combustible liquids
시험 성적서(COA)
제품의 로트/배치 번호를 입력하여 시험 성적서(COA)을 검색하십시오. 로트 및 배치 번호는 제품 라벨에 있는 ‘로트’ 또는 ‘배치’라는 용어 뒤에서 찾을 수 있습니다.
관련 콘텐츠
Learn about biomarkers of cardiovascular disease (CVD) and discover how multiplex CVD panels, such as MILLIPLEX® multiplex assays, are being used to increase the efficiency of biomarker discovery for CVD research.
Uncover tips and tricks to multiplexing in this guide from the experts. Learn how you can enhance the power of your research with MILLIPLEX® multiplexing, including industry guidance for immunoassays, multiplexing tips, one-day MILLIPLEX® multiplex assays for limited lab time, and more.
-
Are the SAA beads in the Human CVD Panel 2 kit designed to target SAA-1?
1 답변-
도움이 되었습니까?
-
-
Can you provide more information about the statement in the Human CVD 2 Milliplex kit (HCVD2MAG-67K) that specifically mentions D-Dimer should be measured in serum, not plasma?
1 답변-
D-Dimer is not present in plasma as it is only found in the blood after a blood clot is degraded. Therefore, D-Dimer is not naturally present in plasma unless the coagulation system has been activated, and it is commonly measured in serum only. In the plasma samples tested in-house, all samples were detectable but at the same concentration, so the data isn't reliable with plasma samples. The kit works well with serum samples, with normal ranges of 142 – 5146 ng/mL and a median of 433 ng/mL.
도움이 되었습니까?
-