テクニック
HPLC: suitable
試験パラメータ
column: Kromasil 5-AmyCoat, 25 cm x 4.6 mm, 5 μm particles (K08670348)
mobile phase: 84:16 hexane: ethanol with 0.1% TFA
flow rate: 1 mL/min
pressure: 1117 psi (77 bar)
column temp.: 25 °C
detector: UV, 254 nm
injection: 20 μL
sample: 50 μg/mL atorva
適合性
application for HPLC
アプリケーション
pharmaceutical (small molecule)
アナリシスノート
Atorvastatin calcium also known as HMG-CoA reductase inhibitors in a class of lipid-lowering medications that reduces levels of “bad” cholesterol (low-density lipoprotein, or LDL) and triglycerides in the blood and raise “good” cholesterol (high-density lipoprotein, or HDL). The enantiomer of atorvastatin (related compound E) could qualitatively and quantitatively differ in their activities, in pharmacokinetic behavior or pharmacodynamics. The application exhibits baseline resolution under 9 minutes of a racemic mixture of atorvastatin calcium and related compound E on a 5.0 μm Kromasil 5-AmyCoat Chiral.
法的情報
Kromasil is a registered trademark of EKA Chemicals AB
アナライト
詳細
Atorvastatin related compound E (PHR1872)
Atorvastatin Calcium, Pharmaceutical Secondary Standard; Certified Reference Material
Choose from one of the most recent versions:
試験成績書(COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact カスタマーサポート
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.