コンテンツへスキップ
Merck

G005887

LC/MS (TOF) Analysis of DL-α-Hydroxyglutaric Acid Enantiomers on Astec® CHIROBIOTIC® R, Fast Separation

application for HPLC

ログイン組織・契約価格を表示する


About This Item

テクニック

HPLC: suitable

試験パラメータ

column: Astec CHIROBIOTIC® R, 10 cm x 4.6 mm I.D., 5μm particles (13022AST)
mobile phase: 5:95, (water:methanol) with 0.3% acetic acid and 0.1% ammonium hydroxide
flow rate: 1 mL/min
pressure: 1700 psi (117 bar)
column temp.: 25 °C
detector: MS, ESI(-) TOF, SIM, m/z 147.03, split 100:1
injection: 10 μL
sample: DL-a-Hydroxyglutaric acid, 1 mg/mL in 50:50, water:methanol

適合性

application for HPLC

アプリケーション

clinical

アナリシスノート

Chiral 2-hydroxyglutarates (2-OHG) are important molecular signatures of both healthy and diseased biological cells, their specific biochemical pathways and inborn errors of metabolism. The chiral differentiation and quantification of D-2-OHG and L-2-OHG is key for characterizing neurometabolic disorders like the 2-hydroxyglutaric acidurias that cause neurological impairment early in life. In this application, an Astec® CHIROBIOTIC® R column run under polar ionic mobile phase conditions achieved a rapid separation of the D and L enantiomers of 2-OHG under conditions amenable for sensitive LC-MS/MS detection.

法的情報

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC

アナライト

詳細

    DL-α-Hydroxyglutaric acid disodium salt, ≥95.0% (GC)

Choose from one of the most recent versions:

試験成績書(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.