コンテンツへスキップ
Merck

1.17144

Millipore

MAS-100 Iso MH® Remote Microbial Air Sampler

Pre-assembled for two heads with USB, digital I/O and ProfiBus interfaces

ログイン組織・契約価格を表示する


About This Item

UNSPSCコード:
41103403
NACRES:
NA.74

物質

aluminum hardware

認証

according to ISO 14698

特徴

installed ( )

包装

pkg of 1 kit

メーカー/製品名

MAS-100 Iso MH®

アプリケーション

air monitoring
pharmaceutical

適合性

for use with 90 mm Petri dishes

保管温度

no temp limit

類似した製品をお探しですか? 訪問 製品比較ガイド

詳細

The MAS-100 ISO MH® control with Profibus, 2 heads is designed to monitor the microbial contamination of the air in isolators and RABS but can also be used as fixed point sampling in cleanrooms. Like all other MAS-100 air samplers, it′s based on the Andersen impaction principle. The control unit with electronic and moving blower parts can be located outside the critical zone, while the single sampling head for standard 90 to 100 mm Petri dishes collect samples in the containment zone. The tubing length can be up to 10 m when used with VHP decontamination or up to 50 m without.
The innovative double valve system enables the sampling heads to be integrated into the decontamination process with VHP. Each unit has both an air and a sterilization pump that operate autonomously.
The fully automatic calibration of the device guarantees maximum efficiency and accuracy. The system measures the air flow and regulates it to a constant value of 100 liters/min. These systems have the highest airflow accuracy of ±2.5%, compared to others that can be as high as ±10%. The instruments are designed and built according to GAMP 5 and are independently validated according to ISO 14698-1 with very good physical efficiency providing a D50 value below 1.1μm and a high biological efficiency result. The efficiencies and the D50 values are independent of the tube length and are comparable to the other members of the MAS-family.
This version is equipped with two valves for two sampling heads and RS232, USB, digital IN/OUT and Profibus interfaces for communication. In case another communication interface is required, MAS-100 ISO MH® Control unit with Ethernet, 2 heads is available.
The type of sample base (standard or easy clean version) as well as the perforated lids can be selected separately according to the specific needs of the sample location and workflow.

アプリケーション

MAS-100 ISO MH® Control unit with Profibus, 2 heads is suitable for microbial monitoring in isolators and RABS. The microbial air sampling devices are extensively used in Pharmaceutical applications.

構成

Control unit equipped with two valve (including cables) for two sampling heads, a range of phoenix connectors, Profibus module, USB cable, digital I/O, USB stick with all manuals and documentation.

法的情報

MAS-100 ISO MH is a registered trademark of MBV AG

付属品

製品番号
詳細
価格

関連製品

製品番号
詳細
価格

保管分類コード

10-13 - German Storage Class 10 to 13


試験成績書(COA)

製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

関連コンテンツ

最も厳しい要求が課される製造現場において、私たちの微生物エアーモニタリングシステムは、非常に高い精度でアイソレーターやRABSにおいて空中浮遊菌をサンプリングおよびモニタリングすることができます。

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)