HPLC: suitable
Parametri analitici
sample/matrix: urine samples spiked at 100 ng/mL with each target analyte (To ensure full ionization of the analytes, spiked samples were treated with formic acid to a final concentration of 0.1% formic acid.)
SPE tube/cartridge: Supel-Select SCX, 30 mg/1 mL (54240-U)
condition: 1 mL 1% formic acid acetonitrile then 1 mL water
sample addition: 1 mL spiked water blank or urine
washing: 1 mL water, 1 mL 1% formic acid acetonitrile, 1 mL water
elution: 2 mL 10% ammonium hydroxide in acetonitrile
eluate post-treatment: thoroughly mix via vortex agitation, evaporate 1 mL aliquot to dryness, reconstitute in 100 μL water:methanol
column: Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
mobile phase: (A) 5 mM ammonium formate acetonitrile; (B) 5 mM ammonium formate water; (98:2, A:B); premixed
flow rate: 0.6 mL/min
pressure: 1842 psi (127 bar)
column temp.: 35 °C
detector: MS, ESI(+), TIC, m/z 100-1000
injection: 1 μL
sample: 200 μg/L ea. in acetonitrile
Compatibilità
application for HPLC
applicazioni
forensics and toxicology
Risultati analitici
Note legali
Analita
Accessorio
Colonna analitica
Colonnina o piastra per SPE
Componente della fase mobile
Standard
Choose from one of the most recent versions:
Certificati d'analisi (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.