G004431
LC/MS Analysis of Clenbuterol In Plasma on Astec® CHIROBIOTIC® T with Phospholipid Removal using HybridSPE® Phospholipid
application for SPE, application for LC-MS
About This Item
tecniche
LC/MS: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with clenbuterol enantiomers at 10 ng/mL
SPE well plate: HybridSPE-Precipitation 96-well Plate, 50 mg/well (575656-U)
sample addition: 100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile. Mix by vortexing the HybridSPE-PPT plate briefly.
elution: apply vacuum
column: CHIROBIOTIC T, 10 cm x 2.1 mm I.D., 5 μm particles (12018AST)
mobile phase: 10 mM ammonium formate in methanol
flow rate: 0.3 mL/min
column temp.: 30 °C
detector: ABI 3200 QT; ESI(+), MRM: 184/104 m/z (phospholipids); 277.2/203.1 m/z (clenbuterol)
injection: 10 μL
Compatibilità
application for LC-MS
application for SPE
applicazioni
clinical
forensics and toxicology
pharmaceutical (small molecule)
Applicazioni
Note legali
Applicazioni
Colonna analitica
Colonnina o piastra per SPE
Standard
Choose from one of the most recent versions:
Certificati d'analisi (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Chromatograms
application for HPLCapplication for HPLCapplication for HPLCapplication for HPLCIl team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.