Passa al contenuto
Merck
Tutte le immagini(1)

Documenti

KHGLA1THH1

Millipore

Capsule Opticap® XLT, membrana Multilayer Durapore® da 0,45/0,22 µm

Opticap® XLT 10, cartridge nominal length 10 in. (25 cm), filter diam. 4.2 in. (10.7 cm)

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

Codice UNSPSC:
23151806

Materiali

polypropylene filter
polypropylene housing
polypropylene support
silicone seal

Livello qualitativo

Conformità normativa

FDA 21CFR177-182 (all component materials)

Sterilità

non-sterile

Compatibilità con la sterilizzazione

autoclavable compatible

Nome Commerciale

Opticap® XLT 10

Caratteristiche

hydrophilic

Produttore/marchio commerciale

Opticap®

Parametri

≤10.8 mL/min air diffusion at 2.75 bar (40 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
1.05 bar max. differential pressure (15 psid) at 80 °C (Forward:)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure

tecniche

sterile filtration: suitable

Larghezza

7.8 in.

Lunghezza nominale della cartuccia

10 in. (25 cm)

device L

37.6 cm (14.8 in.)

device size

10 in.

Diam. filtro

4.2 in. (10.7 cm)

filter filtration area

1.86 m2

Diam. connessione ingresso

5/8 in.

inlet to outlet width

19.8 cm (7.8 in.)

Diam. connessione uscita

5/8 in.

Impurezze

≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

Matrice

Multilayer Durapore®

pore size

0.45/0.22 μm pore size

Punto di bolla

≥3450 mbar (50 psig), air with water at 23 °C

Raccordi

1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(16 mm (5?8 in.) Hose Barb Inlet and Outlet)

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

Device Configuration: Capsule

Confezionamento

Double Easy-Open bag

Nota sulla preparazione

Sterilization Method:
3 autoclave cycles of 60min at 126°C; not in-line steam sterilizable

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).

Risultati analitici

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
TOC/Conductivity: After sterilization and a controlled water flush, samples exhibited less than 500ppb TOC per USP < 643 > and less than 1.3μS/cm per USP < 645 > at 25°C.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 1.5 L

Altre note

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Note legali

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany

Non trovi il prodotto giusto?  

Prova il nostro Motore di ricerca dei prodotti.


Certificati d'analisi (COA)

Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.