LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 2
材料
分析柱
产品编号
说明
价格
應用
標準
产品编号
说明
价格
Butylone hydrochloride solution
1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®Ethylone hydrochloride
1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®3,4-亚甲基二氧吡咯烷酮盐酸盐(MDPV)标准液 CRM 溶液
1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®αα-盐酸吡咯烷基苯戊酮 盐酸盐 溶液
1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®CONDITIONS
column
Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
mobile phase
[A] 5 mM ammonium formate in acetonitrile: [B] water; (98:2, A:B)
flow rate
0.6 mL/min
pressure
267 psi (18.4 bar)
column temp.
35 °C
detector
MS, ESI(+) TOF, Summed Ion Chromatogram
injection
2 μL
sample
500 ng/mL in acetonitrile
说明
分析報告
HILIC mode operation on an Ascentis Express column provided baseline resolution of these bath salts, including isobaric compounds. The highest quality LC-MS solvents were used to supply low background interference and low particulate contaminants for robust; trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.