跳转至内容
Merck

HPLC Analysis of Four Proteins on Zenix® SEC-300, Comparison of Column Internal Diameter

HPLC Analysis of Four Proteins on Zenix® SEC-300, Comparison of Column Internal Diameter application for HPLC

材料

分析色谱柱

产品编号
说明
价格

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 2.1 mm, 3 μm particle size, stainless steel

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

标准

产品编号
说明
价格

肌红蛋白 来源于马心脏

≥90% (SDS-PAGE), essentially salt-free, lyophilized powder

IgG 来源于人类血清

reagent grade, ≥95% (SDS-PAGE), essentially salt-free, lyophilized powder

尿嘧啶

≥99.0%

甲状腺球蛋白 来源于牛甲状腺

(For Gel Filtration Chromatography)

牛血清白蛋白 来源于牛血清

lyophilized powder, ≥96% (agarose gel electrophoresis)

CONDITIONS

column

Zenix SEC-300, 30 cm x 2.1 mm I.D., 3 μm particles (Z777024)

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (Z777028)

column

Zenix SEC-300, 30 cm x 7.8 mm I.D., 3 μm particles (Z777033)

mobile phase

0.15 M potassium phosphate monobasic, pH 7.0 with phosphoric acid

flow rate

15, 72, 207 μL/min (2.1 mm ID, 4.6 mm ID, 7.8 mm ID, respectively)

pressure

220, 240, 350 psi (2.1 mm ID, 4.6 mm ID, 7.8 mm ID, respectively)

column temp.

25 °C

detector

UV, 215 nm

injection

0.3, 1.4, 4.0 μL (2.1 mm ID, 4.6 mm ID, 7.8 mm ID, respectively)

sample

(Waters SEC200 Protein Std)

说明

分析说明

Smaller column internal diameters (ID) are attractive for reasons of low solvent consumption. However, when adequate sample is available, larger ID analytical columns can be advantageous because they provide better column performance.

法律信息

Zenix is a registered trademark of Sepax Technologies