GC Analysis of Nitrogen-Containing Herbicides in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Nitrogen-Containing Herbicides in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

固相微萃取纖維

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

4 mL (water + 100ng/mL each analyte 1g/mL NaCl), pH 2, in 4.6 mL vial

SPME fiber

polyacrylate, 85 μm (57304)

extraction

immersion, ambient temp., 50 min (constant stirring)

desorption process

5 min, 230 °C

column

5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)

oven

40 °C (5 min) to 100 °C at 30 °C/min, to 275 °C at 5 °C/min

inj. temp.

splitless, 230 °C

detector

MS (Ion Trap Manifold: 250°C; scan range m/z=45-400 at 0.6 sec/scan)

carrier gas

helium, 40 cm/sec, set at 40 °C