跳转至内容
Merck

G005681

HPLC Analysis of Bisphenol A, Dimethylbisphenol A and Nonylphenol on Ascentis® Express RP-Amide (UV Detection)

application for HPLC

登录查看公司和协议定价


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis Express RP-Amide, 5 cm x 4.6 mm I.D., 2.7 μm particles (53922-U)
mobile phase: [A] water; [B] acetonitrile
gradient: 50 to 100% B in 10 min
flow rate: 1.0 mL/min
column temp.: 35 °C
detector: UV, 280 nm
injection: 10 μL
sample: 50 mg/L of each in 50:50, water: acetonitrile

適合性

application for HPLC

應用

environmental
food and beverages

分析報告

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of milk thistle related compounds

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Bisphenol A, ≥99%
    Nonylphenol, PESTANAL®, analytical standard, technical mixture

Choose from one of the most recent versions:

分析证书(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门