跳转至内容
Merck
所有图片(1)

主要文件

G005538

GCxGC Analysis of FAMEs in Human Plasma on Equity-1 x SUPELCOWAX 10, Non-Polar to Polar Strategy

suitable for GC

登录查看公司和协议定价


About This Item

技術

gas chromatography (GC): suitable

測試參數

column: (1D) Equity 1, 30 m x 0.25 mm I.D., 0.25 μm (28406-U)
column: (2D) SUPELCOWAX 10, 0.95 m x 0.10 mm I.D., 0.10 μm (24343)
oven: 180 °C, 3 °C/min to 280 °C (10 min)
oven: (modulator) LMCS Everest longitudinally modulated cryogenic system, modulation period 6 sec
inj. temp.: 280 °C
detector: FID, 300 °C, 125Hz; MS, 20000 amu/sec
carrier gas: hydrogen
sample: human plasma

應用

clinical

一般說明

The 0.95 m SUPELCOWAX 10 column was obtained by cutting lengths from a standard length column (e.g. 5, 10, 15 m).

分析報告

Sample prep: 100 mL of plasma was saponified with 1 mL of methanolic sodium methoxide (0.5% w/v) at a temperature of 100 °C for 15 min in a closed Pyrex tube. The subsequent methylic esterification was achieved with 1 mL boron trifluoride-methanol complex at a temperature of 100 °C for 15 min. The fatty acid methyl esters (FAMEs) were extracted by adding 1 mL of n-hexane to the mixture. One mL of a saturated NaCl solution was added to the entire mixture, which was then agitated manually for 2 min, followed by centrifugation at 3000 rpm for 10 min. After, anhydrous Na2SO4 was added to the Pyrex tube for drying and 0.9 mL of the n-hexane FAME layer was transferred to a GC injector vial. A 0.1 mL volume of 1000 ppm internal standard (C13:0) was also added to the GC vial.

法律資訊


历史批次信息供参考:

分析证书(COA)

Lot/Batch Number

抱歉,我们目前尚未线上提供该产品的COA。

如需帮助,请联系 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门