Accéder au contenu
Merck
Toutes les photos(1)

Key Documents

G005721

LC/MS Analysis of Pilocarpine and Betaxolol in the Presence of Benzalkonium Chlorides on Ascentis® Express HILIC

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express HILIC, 15 cm x 4.6 mm I.D., 2.7 μm particles (53981-U)
mobile phase: [A] 2 mM arrmonium acetate, pH 6.7 in [B] 10:90 (v/v) water:acetonitrile
flow rate: 1.0 mL/min
pressure: 1160 psi (80 bar)
column temp.: 35 °C
detector: ESI(+), XlC, m/z 304, 332 (BACs), m/z 209 (pilocarpine), m/z 308 (betaxolol)
injection: 2 μL
sample: 10 μg/mL each in acetonitrile

Adéquation

application for HPLC

Application(s)

pharmaceutical (small molecule)

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Benzalkonium chloride

Colonne analytique

Étalon

Réf. du produit
Description
Tarif

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique