TLC Analysis of Ingredients in Broncholytic Agents on HPTLC Silica Gel 60 RP-18 WF 254s with concentrating zone
Matériaux
À utiliser avec
Réf. du produit
Description
Tarif
Cuve de développement de plaques CCM Aldrich® - rectangulaires
L × H × W 27.0 cm × 26.5 cm × 7.0 cmCuve de développement de plaques CCM Aldrich® - rectangulaires
L × H × W 7.5 cm × 15.5 cm × 8.0 cmSeringues TLC Hamilton®
1702, volume 25 μL, needle size 22s ga, needle L 51 mm (2 in.),coated tip
Seringues TLC Hamilton®
1705, volume 50 μL, needle size 22 ga, needle L 51 mm (2 in.),coated tip
Seringues TLC Hamilton®
1701, volume 10 μL, needle size 26s ga, needle type, fixed, needle L 51 mm (2 in.),coated tip
Cuve de développement de plaques CCM Aldrich® - rectangulaires
L × H × W 17.5 cm × 16.0 cm × 8.2 cmSeringues TLC Hamilton®
1710, volume 100 μL, needle size 22 ga, needle L 51 mm (2 in.),coated tip
CONDITIONS
plate
HPTLC pre-coated plate RP-18 wF 254s with concentrating zone (1.13124.0001)
eluent
[A] acetonitrile; [B] water; (30:70, A:B) containing 0.1 M TEACI
migration distance
7 cm
chamber
normal chamber without chamber saturation
sample
5 μL (10 mg/5 mL)
detector
(in situ evaluation with TLC/HPTLC scanner (Camag) UV 265)