HPLC Analysis of Sugars on Ascentis® Express HILIC (Mobile Phase 5:95)

application for HPLC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

técnicas

HPLC: suitable

parámetros analíticos

column: Ascentis Express HILlC, 15 cm x 4.6 mm I.D., 2.7 μm particles (53981-U)
mobile phase: [A] water; [B] acetonitrile; (5:95, A:B)
flow rate: 1.0 mL/min
column temp.: 35 °C
detector: ELSD, 55 °C, 3.5 bar nitrogen
injection: 10 μL

idoneidad

application for HPLC

aplicaciones

food and beverages

Nota de análisis

This application demonstrates the suitability of the Ascentis Express HILIC for the analysis of sugars. Structures along with the optimized chromatogram obtained on the Ascentis Express HILIC are presented. Separation is by normal-phase mechanism as a decrease in mobile phase water content results in increased retention

Información legal

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

columna analítica

Referencia del producto
Descripción
Precios

patrón


Choose from one of the most recent versions:

Certificados de análisis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico