Conductivity is not measured at different temperatures to create a correction table. Over typical measurement temperatures, the industry standard is a correction factor of 2% per °C. The CRM should be used to calibrate or confirm the conductivity meter, which should read 25 µmhos/cm at 25 °C. If a reading of 25 µmhos/cm is not obtained, the instrument may require adjustment. Alternatively, the standard correction factor or another factor can be used to calculate acceptance criteria at the measurement temperature.
Seleccione un Tamaño
| A ustedes/SKU | Disponibilidad | Precio |
|---|---|---|
500 mL | Disponible para enviar HOYdesdeLyon | 107,00 € |
Acerca de este artículo
grade
certified reference material
Quality Segment
CofA
current certificate can be downloaded
manufacturer/tradename
RTC COND-25-0.500
concentration
in H2O
application(s)
environmental
food and beverages
pharmaceutical
water monitoring
format
single component solution
storage temp.
2-30°C
General description
All information regarding metrological traceability, uncertainty, homogeneity and stability is documented in the certificate of analysis (CoA).
Application
Other Notes
1 of 1
Este artículo | |||
|---|---|---|---|
| description certified reference material | description certified reference material | description certified reference material | description certified reference material |
| application(s) environmental | application(s) environmental | application(s) environmental | application(s) environmental |
| format single component solution | format single component solution | format single component solution | format single component solution |
| grade certified reference material | grade certified reference material | grade certified reference material | grade certified reference material |
| manufacturer/tradename RTC COND-25-0.500 | manufacturer/tradename RTC COND-75-0.500 | manufacturer/tradename RTC COND-147-1 | manufacturer/tradename RTC COND-50-0.500 |
| concentration in H2O | concentration in H2O | concentration in H2O | concentration in H2O |
| storage temp. 2-30°C | storage temp. 2-30°C | storage temp. 2-30°C | storage temp. 2-30°C |
Clase de almacenamiento
10 - Combustible liquids
wgk
nwg
flash_point_f
Not applicable
flash_point_c
Not applicable
Elija entre una de las versiones más recientes:
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
-
when i use this solution, if the temperature value not 25 Celcius,, example the temperature value is 29 celcius, what value will be read?
1 respuesta-