794-0399
GC Analysis of a 20-Component Organochlorine Pesticide Mix in Water on a 5% Phenyl Column after SPE using Supelclean™ ENVI™-8
suitable for GC, application for SPE
Synonym(e):
Chlorinated pesticides in hazardous waste
About This Item
Methode(n)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Testparameter
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: waste water spiked with pesticides
SPE tube/cartridge: Supelclean ENVI-8, glass hardware, 500 mg/6 mL (57107 (500 mg/6 mL); 57108 (1 g/6 mL))
condition: 3 mL methanol; h 2 mL 5% methanol in water
sample addition: load sample at 5 mL/min
drying: 2-3 min with nitrogen stream
elution: 2 x 4 mL hexane:acetone (90:10), drop wise. allow to soak into bed
column: 5% Phenyl Polysiloxane, 15 m × 0.53 mm I.D., 0.50 μm (Supelco equivalent, Equity-5, 28252-U)
oven: 150 °C (0.5 min) to 275 °C (5 min)
inj. temp.: 200 °C
detector: ECD, 300 °C
carrier gas: helium
injection: 1 μL split/splitless (45 sec delay)
Eignung
application for SPE
Anwendung(en)
environmental
food and beverages
Anwendung
Add 5 mL methanol to 100 mL aqueous hazardous waste. Adjust to pH 5-7 if necessary.
Rechtliche Hinweise
Analyt
SPE-Röhrchen oder -Platte
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Dokumente section.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.